How to use Accurex LPM-101 lipid profile meter with LPS-101 test strip showing code chip setup second blood drop 35 microlitre sample and two minute result

How to Use Accurex Lipid Profile Meter & LPS-101 Test Strips: Step-by-Step Guide

The Accurex Lipid Profile Meter / Blood Lipid Analysis System is a portable point-of-care system designed for quantitative lipid testing using compatible dry-chemistry test strips.

When used with the LPS-101 3-in-1 Lipid Panel Strip, one test directly measures:

  • Total Cholesterol – TC
  • HDL Cholesterol
  • Triglycerides – TG

The compatible meter can then calculate:

  • LDL Cholesterol
  • TC/HDL Ratio

Correct testing depends on more than simply applying blood to a strip. The correct code chip, specimen type, strip, sample volume and application timing must all be checked before the test begins.

This guide explains the complete Accurex LPM-101 testing workflow using the LPS-101 lipid profile strip.


Accurex LPS-101 Testing at a Glance

Testing Detail Accurex LPS-101
Compatible Meter Accurex / LPM-101 Blood Lipid Analysis System
Strip LPS-101 3-in-1 Lipid Panel Strip
Directly Measures TC + HDL + TG
Calculated Results LDL + TC/HDL Ratio
Sample Volume Approximately 35 µL
Capillary Sample Use the second blood drop
Test Time Approximately 2 minutes
Operating Temperature Approximately 10°C–35°C
Calibration Matching code-chip system

Before You Start: What You Need

For an LPS-101 lipid profile test, prepare:

  • Accurex LPM-101 Lipid Profile Meter
  • LPS-101 Lipid Profile Test Strip
  • Code chip supplied with the LPS-101 strip box
  • Safety lancet or compatible lancing device
  • Capillary transfer tube or pipette
  • Clean gloves where appropriate
  • Sharps / biomedical-waste disposal container

Do not begin the test until the required supplies are ready.


Step 1: Check the Accurex Meter

Before testing, inspect the meter.

Confirm that:

  • The meter is clean.
  • The test-strip holder is free from contamination.
  • The meter has adequate battery charge.
  • The meter is not being charged during testing.
  • The meter and test materials are within the permitted operating temperature.

The manufacturer specifies testing conditions of approximately 10°C–35°C.

If the meter has recently been moved from a very hot or cold environment, allow it and the testing materials to reach suitable operating conditions before testing.


Step 2: Check the LPS-101 Test Strip Pack

Before removing a strip, check:

  • The product is LPS-101.
  • The strip pack is intended for the compatible Accurex / LPM-101 system.
  • The printed expiry date has not passed.
  • The packaging is not damaged.
  • The strips have been stored correctly.

Do not use a strip that is expired, wet, discoloured, bent or visibly damaged.

View Accurex Lipid Profile Test Strips – 25 Pack


Step 3: Insert the Correct Code Chip

Each new box of Accurex test strips is supplied with a corresponding code chip.

Before testing:

  1. Locate the code chip supplied with the LPS-101 strip box.
  2. Check the code printed on the chip.
  3. Compare it with the code printed on the strip box, canister or foil packaging.
  4. Insert the matching code chip into the meter's code-chip slot.
  5. Turn on the meter.
  6. Verify that the displayed code matches the strip-pack code.

Do not proceed if the codes do not match.


Why Code-Chip Matching Matters

The code chip provides calibration information corresponding to the strip lot.

Using:

  • An old chip
  • A chip from another strip box
  • A chip from TCS-101, TGS-101 or HLS-101
  • A chip from a different lot

should not be assumed to provide the correct calibration for the current LPS-101 strips.

Keep each strip box and its code chip together.


Step 4: Confirm the Specimen Type

The Accurex meter supports different specimen types, including:

  • Capillary whole blood
  • Venous whole blood
  • Plasma
  • Serum

Before testing, verify that the specimen setting displayed by the meter corresponds to the sample you are actually using.

If it does not, select the correct specimen type according to the meter instructions before continuing.


Step 5: Insert the LPS-101 Strip

Wait until the meter is ready for a test strip.

Then:

  1. Remove one LPS-101 strip from its protective packaging.
  2. Hold the strip by the appropriate handling area.
  3. Follow the arrows printed on the strip.
  4. Insert the strip into the meter's test-strip holder.
  5. Push it completely into the holder until it cannot be inserted further.

The meter will prepare for specimen application.


Step 6: Wait for the Blood-Drop Symbol

Do not apply the specimen immediately after inserting the strip.

Wait until the meter displays the flashing:

Blood-drop / specimen symbol

This indicates that the system is ready to receive the sample.


Step 7: Prepare for Capillary Blood Collection

For fingertip testing:

  1. Wash the hands with soap and water.
  2. Dry the hands thoroughly.
  3. Prepare a sterile safety lancet or lancing device.
  4. Select an appropriate fingertip puncture site.
  5. Perform the finger prick.

Do not test from a finger that is still wet from washing, alcohol or another liquid.


Step 8: Wipe Away the First Drop of Blood

For capillary testing, the manufacturer's LPM-101 procedure specifically instructs users to:

Wipe away the first drop of blood.

Then allow a second drop to form.

The second drop is used for the lipid test.


Step 9: Collect Approximately 35 µL of Blood

The LPS-101 3-in-1 lipid panel requires approximately:

35 µL of specimen

The manufacturer recommends using a:

  • Capillary transfer tube
  • or suitable pipette

to obtain the correct volume.

When using the capillary transfer tube:

  1. Hold the tube slightly downward.
  2. Touch the tube tip gently to the blood drop.
  3. Allow capillary action to draw the sample.
  4. Continue until the specimen reaches the indicated fill line.

Do not squeeze the capillary tube while collecting the blood sample.


Why Sample Volume Matters

The LPS-101 strip simultaneously measures multiple lipid parameters.

An inadequate specimen can affect how the test strip fills and reacts.

For reliable testing:

  • Collect the required quantity.
  • Avoid partial filling.
  • Do not guess the volume.
  • Use the recommended transfer device.

Step 10: Apply the 35 µL Sample to the Strip

While the blood-drop symbol is flashing:

  1. Position the transfer tube or pipette over the specimen application area.
  2. Align it with the centre of the designated application region.
  3. Apply the approximately 35 µL sample.
  4. Avoid touching or scratching the strip with the tube or pipette tip.

The specimen should be applied cleanly to the designated area.


Do Not Apply Blood Too Early

Only apply the specimen after the meter shows that it is ready.

Applying blood before the appropriate prompt can interfere with the normal testing workflow.


Do Not Add More Blood Later

If an inadequate amount of specimen was applied and the test fails, do not keep adding blood to the same strip.

Instead:

  1. Remove and discard the used strip.
  2. Insert a new LPS-101 strip.
  3. Collect the correct amount of fresh specimen.
  4. Repeat the test.

Step 11: Wait While the Accurex Meter Performs the Test

After sufficient specimen has been applied, the meter begins testing automatically.

The display shows flashing dashes while the test is in progress.

Do not:

  • Remove the test strip
  • Switch off the meter
  • Move the strip
  • Add more sample

during the active measurement.


Step 12: Read the Results After Approximately 2 Minutes

The Accurex lipid profile results are displayed in approximately:

2 minutes

With the LPS-101 lipid panel, the system can provide:

  • Total Cholesterol – TC
  • HDL Cholesterol
  • Triglycerides – TG
  • Calculated LDL Cholesterol
  • Calculated TC/HDL Ratio

Measured vs Calculated Accurex Results

Parameter Result Type
Total Cholesterol Measured
HDL Cholesterol Measured
Triglycerides Measured
LDL Cholesterol Calculated
TC/HDL Ratio Calculated

Step 13: Remove the Used Test Strip

Once the result has been displayed:

  1. Record or review the result as required.
  2. Remove the used test strip.
  3. Dispose of the strip appropriately.
  4. Dispose of the lancet in a suitable sharps container.

Test strips and lancets are single-use items.


Do Not Reuse LPS-101 Test Strips

Each LPS-101 strip is intended for one test only.

Never:

  • Wash a used strip
  • Reuse it with a new sample
  • Use the same strip for another patient
  • Return a used strip to its packaging

Step 14: Clean the Meter After the Testing Session

The manufacturer recommends cleaning the Accurex / LPM-101 meter after each day of testing.

General cleaning includes:

  • Wiping the meter surface with a clean cotton cloth
  • Using a slightly damp cloth where necessary
  • Using a dry soft cloth for the LCD and sensor area
  • Keeping the strip holder clean

Avoid allowing liquid to enter:

  • The test-strip holder
  • The code-chip slot
  • The USB port

Do Not Scratch the Optical Sensor Window

The meter relies on an optical system to read the strip reaction.

When cleaning the sensor region:

  • Use a soft cotton swab where appropriate.
  • Do not use sharp objects.
  • Do not scratch the transparent sensor window.
  • Do not use aggressive organic solvents.

Can the Accurex Meter Be Used While Charging?

No.

The LPM-101 uses a rechargeable lithium battery, but the meter is not intended to perform testing while actively charging.

Charge the device before beginning a testing session if battery power is low.


Using Venous Whole Blood

The Accurex system can also support venous whole blood testing.

For professional venous specimens:

  • Use the anticoagulant specified by the manufacturer.
  • Mix whole-blood specimens adequately before testing.
  • Use the required sample volume.
  • Apply the specimen with an appropriate pipette or transfer tube.
  • Do not touch the strip with the pipette tip.

Follow the current LPS-101 package insert for specimen handling and testing timelines.


Using Serum or Plasma

The Accurex Blood Lipid Analysis System also supports serum and plasma specimens according to the manufacturer's instructions.

Before testing:

  • Select the correct specimen type on the meter.
  • Ensure proper sample preparation.
  • Use an appropriate transfer device.
  • Apply the required sample volume.

Professional users should follow their laboratory's applicable specimen-handling procedures.


Capillary vs Venous Blood: Do Not Forget the Meter Setting

Because the system supports different specimen types, the meter should be configured correctly before testing.

Do not assume that the previous sample-type setting is correct for the next patient or specimen.

Check it before every testing session.


What If the Meter Does Not Recognise the Strip?

Check:

  • That LPS-101 is compatible with the meter
  • The strip orientation
  • Whether the strip is completely inserted
  • Strip expiry
  • Physical damage
  • The test-strip holder for contamination

Do not force the strip into the meter.


What If the Meter Shows a Code-Chip Error?

The LPM-101 manual identifies code-chip recognition as an essential part of testing.

If the chip is missing, removed or incorrect:

  • Check the code chip.
  • Confirm it belongs to the current strip box.
  • Reinsert it correctly.
  • Verify that the displayed code matches.

Do not continue testing with a known code mismatch.


What If the Meter Shows an Expired-Strip Error?

Check:

  • The printed strip expiry date
  • The meter date setting
  • The strip packaging
  • The correct code chip

Use a fresh, in-date strip where required.


What If the Temperature Is Too High or Too Low?

The LPM-101 operates within approximately:

10°C–35°C

If the surrounding temperature is outside that range, the meter may prevent testing or display an environmental-temperature error.

Move the meter and testing materials to an appropriate environment and allow them to equilibrate before repeating the test.


What If the Result Seems Unexpected?

Review the complete testing process:

  • Correct strip used?
  • Strip within expiry?
  • Correct code chip?
  • Correct specimen type selected?
  • Correct sample volume?
  • Second blood drop used for capillary testing?
  • Sample applied only after the prompt?
  • Operating temperature correct?
  • Meter and sensor clean?

Where appropriate, repeat using a new strip and fresh specimen.

Unexpected or clinically important results should be confirmed using appropriate laboratory testing when indicated.


Common Accurex Testing Mistakes to Avoid

Mistake Correct Practice
Using the wrong code chip Use the chip supplied with the current strip box
Using the first capillary blood drop Wipe it away and use the second drop
Guessing sample volume Collect approximately 35 µL with an appropriate transfer device
Applying blood before the prompt Wait for the flashing blood-drop symbol
Adding more blood after a failed fill Repeat with a new strip
Removing strip during testing Leave the strip inserted until the result appears
Testing while charging Disconnect charger and test using battery power

Quick Accurex LPS-101 Testing Checklist

  • ✓ Confirm LPS-101 strip compatibility
  • ✓ Check expiry date
  • ✓ Match the code chip
  • ✓ Confirm specimen type
  • ✓ Insert strip completely
  • ✓ Wait for the blood-drop symbol
  • ✓ For capillary blood, wipe away the first drop
  • ✓ Use the second blood drop
  • ✓ Collect approximately 35 µL
  • ✓ Apply sample to the designated application area
  • ✓ Do not add more specimen later
  • ✓ Wait approximately 2 minutes
  • ✓ Dispose of strip and lancet safely
  • ✓ Clean the meter after the day's testing

Frequently Asked Questions

How do I use the Accurex Lipid Profile Meter?

Insert the correct code chip, verify the specimen type, insert an LPS-101 strip, wait for the blood-drop symbol, apply approximately 35 µL of specimen and allow the meter to complete the test.

How much blood does the Accurex LPS-101 strip require?

Approximately 35 µL.

Should I use the first or second blood drop?

For capillary testing, the manufacturer instructs users to wipe away the first drop and use the second blood drop.

How quickly are results available?

Results are displayed in approximately 2 minutes.

What does LPS-101 measure?

It directly measures Total Cholesterol, HDL and Triglycerides. The meter can then calculate LDL and TC/HDL ratio where applicable.

Do I need a code chip?

Yes. Use the code chip supplied with the current strip box and verify that the codes match.

Can I use an old code chip with a new strip box?

Do not assume that it is compatible. Use the chip supplied with each new box of strips.

Can I add more blood if I did not apply enough?

No. If a test fails because of insufficient specimen, repeat using a new strip and the correct amount of fresh specimen.

Can I test while the Accurex meter is charging?

No. The LPM-101 is not operational for testing while charging.

Can I use venous blood?

Yes. Venous whole blood is supported according to the manufacturer's specimen instructions.

Can I use serum or plasma?

Yes. The system supports serum and plasma as well as whole-blood specimens.

What temperature should I test at?

The manufacturer specifies approximately 10°C–35°C for system operation.

Can I reuse an LPS-101 strip?

No. Accurex lipid test strips are single-use only.

How often should the meter be cleaned?

The manufacturer recommends cleaning the meter after each day of testing.

What should I do if the result looks wrong?

Review strip compatibility, expiry, code-chip matching, specimen type, sample volume, collection technique, operating temperature and meter cleanliness. Repeat with a new strip when appropriate and obtain laboratory confirmation for clinically important unexpected results.


Related Accurex Knowledge Hub Guides

Accurex Lipid Profile Meter & Test Strips Complete Guide

Learn about the meter's specifications, lipid parameters, sample types and compatible consumables.

Read the Accurex Complete Guide

Accurex Lipid Profile Test Strip Compatibility Guide

Compare LPS-101, TCS-101, TGS-101 and HLS-101 before buying replacement strips.

Read the Accurex Strip Compatibility Guide

QuikCheck vs Accurex Complete Comparison

Understand the difference between a Total Cholesterol monitor and the multi-parameter Accurex Lipid Profile system.

Read QuikCheck vs Accurex Comparison


Shop Accurex Lipid Profile Products

Accurex Lipid Profile Meter

Reusable portable Blood Lipid Analysis System compatible with the Accurex lipid-strip family.

View Accurex Lipid Profile Meter

Accurex Lipid Profile Test Strips – 25 Pack

LPS-101 3-in-1 compatible lipid-panel strips for TC, HDL and TG measurement with calculated LDL and TC/HDL ratio.

View Accurex Lipid Profile Test Strips – 25 Pack


Explore the Microsidd Knowledge Hub

Read more practical guides on diagnostic meters, test-strip compatibility, point-of-care testing and laboratory diagnostics:

Visit the Microsidd Knowledge Hub


Important Medical & Testing Note

The Accurex Lipid Profile Meter is an in-vitro diagnostic point-of-care system and should be used only with compatible test strips and according to the current manufacturer instructions.

Before every test, verify:

  • The exact strip reference
  • The correct code chip
  • The selected specimen type
  • The strip expiry date
  • The required specimen volume
  • The operating conditions
  • The condition and cleanliness of the meter

Unexpected, abnormal or clinically significant lipid results should be interpreted in the appropriate clinical context and confirmed with laboratory testing when indicated.

These diagnostic products are non-returnable and non-exchangeable.
Please verify the meter model, strip compatibility, required test parameters and pack size before placing your order.