Oxygen inlet pressure range |
276 to 600 kPa (40 to 87 psig)
|
Weight with optional battery |
10.9 kg (24 lb)
|
Weight without optional battery |
10.0 kg (22 lb)
|
Dimensions |
33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D)
|
Suitability |
Pediatric (≥ 20 kg) and Adult
|
AC voltage |
100 to 240 VAC
|
AC frequency |
50 to 60 Hz
|
AC power |
300 V·A
|
Battery operating time |
6 hours in normal conditions
|
C-Flex |
Off, 1 to 3
|
IPAP |
4 to 40 cm H₂O
|
CPAP |
4 to 25 cm H₂O
|
EPAP |
4 to 25 cm H₂O
|
Max P (PPV maximum pressure limit) |
5 to 40 cm H₂O
|
IPAP |
4 to 40 cm H₂O
|
Max P (AVAPS maximum IPAP) |
6 to 40 cm H₂O
|
Inspiratory time |
0.30 to 3.00 sec
|
Max P (AVAPS maximum IPAP) |
6 to 40 cm H₂O
|
Min P (AVAPS minimum IPAP) |
5 to 30 cm H₂O
|
Min P (AVAPS minimum IPAP) |
5 to 30 cm H₂O
|
Oxygen percentage |
21 to 100 %
|
Ramp time |
Off, 5 to 45 min
|
Respiratory rate |
4 to 60 beats per minute
|
Rise time |
1 to 5
|
Triggering and cycling |
Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6
|
AVAPS target tidal volume |
200 to 2000 ml BTPS
|
Max E |
0 to 100 cm H2O/l
|
Max E |
0 to 100 cm H₂O/l
|
Max R |
0 to 50 cm H2O/l/s
|
Max R |
0 to 50 cm H₂Ol/s
|
PPV% |
0 to 100 %
|
Max P (PPV maximum pressure limit) |
5 to 40 cm H₂O
|
Max V (PPV maximum volume limit) |
200 to 3500 ml
|
Breath phase/trigger indicator |
Spontaneous, Timed, Exhale
|
PIP |
0 to 50 cm H₂O
|
PIP |
0 to 50 cm H₂O
|
Patient/total leak |
0 to 200 l/min BTPS
|
Patient Trigger |
0 to 100 %
|
Respiratory Rate |
0 to 90 beats per minute
|
Ti/Ttot |
0 to 91 %
|
Minute volume |
0 to 99.0 l/min BTPS
|
Tidal volume |
0 to 3000 ml BTPS
|
Pressure waveform |
0 - 50 cm H₂O
|
Flow waveform |
240-2401 /min BTPS
|
Volume waveform |
0 - 3,500 ml BTPS
|
Hi Rate (hi respiratory rate alarm) |
5 - 90 beats per minute
|
Lo Rate (low respiratory rate alarm) |
1 - 89 beats per minute
|
Hi Vr (high tidal volume alarm) |
200 - 3,500 ml
|
Lo Vr (low tidal volume alarm) |
OFF - 1,500 ml
|
HIP (high inspiratory pressure alarm) |
5 - 50 cm H₂O
|
LIP (low inspiratory pressure alarm) |
OFF, 1 - 40 cm H₂O
|
Lo VE (low minute ventilation alarm) |
OFF, 0.1 - 991 /min
|
LIP T (low inspiratory pressure delay time) |
5 - 60 sec
|